FTIR光谱仪中傅里叶插值采样方法的研究
Fourier interpolative sampling algorithm in Fourier transform infrared spectrometer
查看参考文献19篇
文摘
|
为解决传统“Brault”采样方法中参考信号过零点计算复杂、耗时较多的问题,提出了基于傅里叶插值技术寻找过零点的方法。通过与其他插值方法进行比较研究,结果表明,这种方法确保了过零点信息准确性的同时简化了数据处理的复杂度,得到的过零点信息线性拟合系数大于0.999。在2 100~2 200 cm~(-1)波段范围内,当参考激光信号误差较小时,傅里叶插值方法得到的仪器SNR是三次样条插值法所得到的1.03倍,而线性插值方法与傅里叶插值方法得到的结果一致;当参考激光信号误差较大时,傅里叶插值方法得到的仪器SNR是线性插值方法得到的仪器SNR的1.05倍。 |
其他语种文摘
|
In order to solve the problem of complex and time-consumption of reference signal zero crossings in the traditional "Brault" sampling method, a method based on Fourier interpolation technique was proposed to find the zero-crossing. Compared with other interpolation methods, the results showed that this method could ensure the accuracy of zero-crossing information and simplify the complexity of data processing. The linear fitting coefficient of zero-crossing information obtained was greater than 0.999. In the range of 2 100-2 200 cm~(-1), when the error of reference laser signal was small, the instrumental SNR obtained by the Fourier interpolation method was 1.03 times that obtained by the cubic spline interpolation method, and the result obtained by the linear interpolation method was consistent with Fourier interpolation method. When the error of reference laser signal was relatively large, the instrumental SNR obtained by the Fourier interpolation method was 1.05 times of the SNR obtained by the linear interpolation method. |
来源
|
红外与激光工程
,2018,47(1):0123001-1-0123001-6 【核心库】
|
DOI
|
10.3788/IRLA201847.0123001
|
关键词
|
采样方法
;
傅里叶变换红外光谱仪
;
“Brault”方法
;
傅里叶插值
;
过采样
|
地址
|
1.
中国科学院安徽光学精密机械研究所, 中国科学院环境光学与技术重点实验室, 安徽, 合肥, 230031
2.
中国科学技术大学, 安徽, 合肥, 230026
|
语种
|
中文 |
文献类型
|
研究性论文 |
ISSN
|
1007-2276 |
学科
|
物理学;自动化技术、计算机技术 |
基金
|
国家自然科学基金
|
文献收藏号
|
CSCD:6178106
|
参考文献 共
19
共1页
|
1.
Li Xiangxian. Study on temperature dependence of the greenhouse gases and carbon isotope ratio spectral analysis.
Infrared and Laser Engineering. (in Chinese),2015,44(4):1178-1185
|
被引
1
次
|
|
|
|
2.
Tong Jingjing. Measurement and study of 1,3-butadiene based on open path FTIR spectroscopy.
Infrared and Laser Engineering. (in Chinese),2013,42(1):239-243
|
被引
1
次
|
|
|
|
3.
Li Yan. Based on trigger sampling method and phase correction of infrared spectrum measurement applications.
Spectroscopy and Spectral Analysis. (in Chinese),2015,35(7):2054-2059
|
被引
1
次
|
|
|
|
4.
Saptari V. Fourier-Transform Spectroscopy Instrumentation Engineering.
Tutorial Texts in Optical Engineering volume TT61,2004
|
被引
1
次
|
|
|
|
5.
Minami K. Zero-crossing sampling of Fourier-transform interferograms and spectrum reconstruction using the real-zero interpolation method.
Applied Optics,1992,31(29):6322-6327
|
被引
2
次
|
|
|
|
6.
Li Zhigang. Data sampling of fourier transform spectrometer interferogram using laser interference fringe.
Journal of Optoelectronics Laser. (in Chinese),2000,11(5):503-506
|
被引
1
次
|
|
|
|
7.
Brault J W. New approach to high-precision Fourier transform spectrometer design.
Applied Optics,1996,35(16):2891-2896
|
被引
6
次
|
|
|
|
8.
Brasunas J C. Uniform time-sampling Fourier transform spectroscopy.
Applied Optics,1997,36(10):2206-2210
|
被引
3
次
|
|
|
|
9.
Pougatchev N S. Advanced technologies high resolution Fourier transform spectrometer for atmospheric studies.
Proceeding of 2000 IEEE,2000:237-243
|
被引
1
次
|
|
|
|
10.
Davis S P.
Fourier transform spectrometry,2001
|
被引
4
次
|
|
|
|
11.
Palchetti L. Spectral noise due to sampling errors in fourier-transform spectroscopy.
Applied Optics,2001,40(19):3235-3243
|
被引
5
次
|
|
|
|
12.
Griffiths P R.
Fourier Transform Infrared Spectrometry. 2nd ed,2007
|
被引
4
次
|
|
|
|
13.
Ren Libing. Digital filter method of oversampling Fourier transform infrared spectrometer.
Infrared and Laser Engineering. (in Chinese),2012,41(6):1438-1441
|
被引
1
次
|
|
|
|
14.
Turner A.
Digitisation of interferograms in Fourier Transform Spectroscopy: US, patent5914780,1999
|
被引
1
次
|
|
|
|
15.
Alber G M. Effect of sampling rate on Fourier transform spectra: oversampling is overrated.
Applied Spectroscopy,1990,44(7):1111-1116
|
被引
2
次
|
|
|
|
16.
Ren Libing. Study on the application of soft zero-crossing detection in the infrared spectral measurement system.
Proceeding of the Chinese Optical Society. (in Chinese),2010
|
被引
1
次
|
|
|
|
17.
Zhou Feng. Timedomain linear interpolation algorithm and its error analysis for estimating signal period.
Chinese Journal of Scientific Instrument. (in Chinese),2011,32(8):1725-1730
|
被引
2
次
|
|
|
|
18.
Wu Jie. Study on harmonic analysis based on cubic spline interpolated arithmetic fourier transform.
Electrical Measurement & Instrumentation. (in Chinese),2016,53(1):15-31
|
被引
1
次
|
|
|
|
19.
Li Xinfu. Comparison of the accuracy for fractal and lagrange interpolation.
Journal of Natural Science of Heilongjiang University. (in Chinese),2008,25(3):323-325
|
被引
2
次
|
|
|
|
|