帮助 关于我们

返回检索结果

基于积分球光源的大动态范围辐射标准传递技术研究
Large Dynamic Range Radiation Standard Transmission Research Based on Integrating Sphere Source

查看参考文献23篇

刘辉 1,2   陈洪耀 1   李佳伟 1   包诗薇 1,2   杨宝云 1   张黎明 1 *  
文摘 为了拓展目前实验室辐射标准的动态范围,提出了基于积分球光源的大动态范围辐射标准传递技术.即在等色温调节的前提下,结合可调光阑和基于大动态范围响应线性硅探测器的监视辐射计将辐照度标准灯-漫射板系统的光谱辐亮度工作标准传递到大动态范围可调节积分球光源各个亮度等级,从而实现辐射标准的动态范围拓展.通过研制大动态范围多级可调节光阑并分析探讨光谱匹配、硅探测器的响应线性对辐射标准传递精度的影响,最终在400~2 500 nm范围内,积分球光源在6个量级内的辐射标准不确定度有望达到4.3%~6.1%,这对光辐射测量定量化发展具有重要的意义.
其他语种文摘 Due to the wide spectrum range and the good performance of exit surface uniformity and angle uniformity,integrating sphere is close to the ideal Lambert uniform light source.In this paper,the principle of large dynamic range radiation standard transmission based on integrating sphere light source was proposed.That is to say,under the premise of equal color temperature adjustment,a monitoring radiometer based on a linear silicon detector with a large dynamic range response,and an adjustable aperture can transfer the spectral radiance working standards of the irradiance standard lamp-solar diffuser system to a large dynamic range adjustable integral sphere light source.Thus,the dynamic range of radiation standards can be expanded.Herein,we prepared a large dynamic adjustable aperture and further discussed the spectral matching and the impact of silicon detector nonlinearity on the accuracy of radiation standard transmission.Finally,in spectrum range of 400~2 500 nm,the uncertainty of 6 orders of magnitude radiation standard is 4.3% ~6.1% in integrating sphere source.It is of great significance to the quantitative development of radiometric.
来源 光子学报 ,2018,47(8):0804002-1-0804002-8 【核心库】
DOI 10.3788/gzxb20184708.0804002
关键词 辐射定标 ; 积分球光源 ; 线性定标 ; 可调光阑 ; 不确定度
地址

1. 中国科学院安徽光学精密机械研究所, 中国科学院通用光学与表征技术重点实验室, 合肥, 230031  

2. 中国科学技术大学, 合肥, 230026

语种 中文
文献类型 研究性论文
ISSN 1004-4213
学科 物理学
基金 中科院科研装备项目 ;  高分专项
文献收藏号 CSCD:6324421

参考文献 共 23 共2页

1.  张鹏. 晨昏轨道卫星的气象应用需求分析. 气象科技进展,2015,5(2):6-12 被引 7    
2.  杨眉. DMSP/OLS夜间灯光数据应用研究综述. 遥感技术与应用,2011,26(1):45-51 被引 43    
3.  王鹤饶. DMSP/OLS数据应用研究综述. 地理科学进展,2012,31(1):11-19 被引 42    
4.  Cao Changyong. Detecting light outages after severe storms using the S-NNP/VIIRS Day/Night band radiances. IEEE Geoscience and Remote Sensing Letters,2013,10(6):1582-1586 被引 7    
5.  吴建生. 基于DMSP/OLS夜间灯光数据的1995-2009年中国地级市能源消费动态. 地理研究,2014,33(4):625-634 被引 1    
6.  Cao Changyong. Quantitative analysis of VIIRS DNB nightlight point source for light power estimation and stability monitoring. Remote Sensing,2014,6:11915-11935 被引 6    
7.  Straka W C. Utilization of the suomi national polar-orbiting parnership(NNP) visible infrared imaging radiometer suite(VIIRS)Day/Night band for arctic ship tracking and fisheries management. Remote Sensing,2017,7:971-989 被引 1    
8.  Hoytc C. Cryogenic radiometer and their application to metrology. Metrologia,1991,28:163-167 被引 14    
9.  Gentile T R. National institute of standards and technology high-accuracy cryogenic radiometer. Applied Optics,1996,35(7):1056-1068 被引 31    
10.  郑小兵. 高精度光辐射定标和标准传递方法. 科学通报,2000,45(12):1341-1344 被引 19    
11.  李建军. 硅陷阱探测器在350-1064nm波段的绝对光谱响应度定标. 物理学报,2009,58(9):6273-6278 被引 1    
12.  陈洪耀. 系统级定标方法实现基于低温辐射计的标准灯400~900nm波段光谱辐照度. 光学学报,2010,30(11):3349-3353 被引 5    
13.  Liao L B. Suomi NPP VIIRS day-night band on-orbit performance. Journal of Geophysical Research:Atmospheres,2013,118:12705-12718 被引 12    
14.  Lee Shih Yan. An overview of NASA VCST SNPP VIIRS day-night band on-orbit calibration methodology. SPIE. 9218,2014:801-815 被引 1    
15.  Ambler T. Beamcon III,a linearity measurement instrument for optical detectors. Journal of Research of the National Institute of Standards and Technology,1994,99(6):751-755 被引 1    
16.  刘辉. 基于双LED光源积分球的硅探测器宽动态范围响应线性定标. 光学学报,2017,37(9):0904001 被引 2    
17.  Sadler F B. Facility for optical calibration at low level light(FOCAL3). Proceedings of 33rd Annual European Meeting on Atmospheric Studies by Optical Methods. 292,2008:87-91 被引 1    
18.  李葆勇. 一种空间相机辐射定标用大面积辐射光源的研制. 光电子·激光,2013,24(3):464-469 被引 4    
19.  袁银鳞. 大孔径可调光谱积分球参考光源研制和检测. 光学学报,2013,33(7):0712004 被引 1    
20.  孙景旭. 4 m直径均匀扩展定标光源. 中国光学,2015,8(5):823-832 被引 5    
引证文献 2

1 王少水 一种亮度可调积分球光源及其应用技术研究 应用光学,2020,41(3):586-590
被引 0 次

2 李佳伟 紫外波段大口径积分球辐射源研制 光学学报,2021,41(20):2012005
被引 0 次

显示所有2篇文献

论文科学数据集
PlumX Metrics
相关文献

 作者相关
 关键词相关
 参考文献相关

版权所有 ©2008 中国科学院文献情报中心 制作维护:中国科学院文献情报中心
地址:北京中关村北四环西路33号 邮政编码:100190 联系电话:(010)82627496 E-mail:cscd@mail.las.ac.cn 京ICP备05002861号-4 | 京公网安备11010802043238号