帮助 关于我们

返回检索结果

基于SIMION软件的空间等离子体探测器的数值仿真
Numerical Simulation Analysis of Space Plasma Detector Based on SIMION

查看参考文献8篇

文摘 空间等离子体探测器采用带偏转板的柱形静电分析器作为传感器,具有大于320°探测视场,可探测约10eV~30keV的热等离子体。利用SIMION软件采用均匀随机抽样建立粒子源的方法仿真探测器的基本特性参数:静电分析器因子、能量分辨率、偏转板因子和偏转角。仿真结果与实测结果吻合很好。静电分析器因子的仿真与实测结果偏差为3.2%,能量分辨率仿真与实测结果偏差为1.5%,偏转角-偏转板因子的关系式与实测结果的相关系数为0.999 4。仿真结果可以有效地应用于该类仪器的设计和定标过程中。
其他语种文摘 The space plasma detector is a kind of cylindrical electrostatic analyzer with the field-of-view larger than 320° and the capability of measuring the plasma of 10 eV~30 keV.The basic parameters of the plasma detector,such as the electrostatic analyzer factor,energy resolution and the relationship between the deflection factor and the deflection angle,were simulated by the software SIMION.The particle source for the simulation was constructed by the method of uniform random sample.The simulation results fit well with the calibration results.The error between the simulation and the calibration on the parameters of electrostatic factor and energy resolution are 3.2% and 1.5%,respectively.The correlation coefficient between the simulation and the calibration on the relationship of deflection angle and deflection factor is 0.999 4.The simulation results show that it can be applied well in the process of plasma instrument design and calibration.
来源 中国空间科学技术 ,2012,32(4):71-76 【核心库】
DOI 10.3780/j.issn.1000-758x.2012.04.010
关键词 空间等离子体探测器 ; 柱形静电分析器 ; 数值仿真
地址

中国科学院空间科学与应用研究中心, 北京, 100190

语种 中文
ISSN 1000-758X
学科 航天(宇宙航行)
基金 国家863计划
文献收藏号 CSCD:4621995

参考文献 共 8 共1页

1.  Martin Wuest. Calibration of Particle Instruments in space physics,2007:50-52 被引 1    
2.  Hardy D A. Precipitating electron and ion detectors (SSJ/4) for the block 5D/flights 6-10 DMSP satellites: calibration and data presentation. Environmental Research Papers, AFGL-TR-84-0317-902, Air Force Geophysics Laboratory, Hanscom AFB,1984 被引 1    
3.  Mccann D. Miniature ion mass analyzer. Planetary and Space Science,2007,55(9):1190-1196 被引 3    
4.  Theodoridis G C. Charged particle transmission through cylindrical plate electrostatic analyzers. Review of Scientific Instruments,1968,39:326-330 被引 1    
5.  Bryce P. The 127° electrostatic analyzer performance as a spectrometer. Canadian Journal of Physics,1973,51:574 被引 1    
6.  Johnstone A D. The geometric factor of a cylindrical plate electrostatic analyzer. Review of Scientific Instruments,1972,43(7):1030-1040 被引 1    
7.  于德江. 圆柱形静电分析器的数值模拟. 空间科学学报,1999,19(1):61-68 被引 2    
8.  Bundaleski N. Optical properties of the 127° cylindrical energy analyzer used in LEIS experiments. Nuclear Instruments and Methods in Physics Research B,2002,198:208-219 被引 1    
引证文献 3

1 李莉莉 MPCVD法合成硼掺杂金刚石薄膜的次级发射性能 中国空间科学技术,2017,37(2):48-53
被引 0 次

2 苏斌 一种2π视场高分辨热等离子体分析仪设计与仿真 宇航学报,2019,40(5):604-610
被引 0 次

显示所有3篇文献

论文科学数据集
PlumX Metrics
相关文献

 作者相关
 关键词相关
 参考文献相关

版权所有 ©2008 中国科学院文献情报中心 制作维护:中国科学院文献情报中心
地址:北京中关村北四环西路33号 邮政编码:100190 联系电话:(010)82627496 E-mail:cscd@mail.las.ac.cn 京ICP备05002861号-4 | 京公网安备11010802043238号