文摘 采用原子力显微镜测量样品表面形貌时,针尖与样品间的相互作用力会使样品表面不同力学性质的区域变形量不同,从而使得形貌测量结果产生偏差.首先研究了引起形貌测量偏差的机理,进一步通过数值计算发现通过引入高频信号与低频信号叠加作为新的形貌测量信号可以近似避免这一类形貌测量偏差.但是对于普通的矩形截面悬臂梁提取高阶信号是比较困难的,通过外加弹簧的结构设计来调节悬臂梁各阶本征频率相对值,使得在实验测量过程中可以较方便地提取悬臂梁振动的高频信号
其他语种文摘 In measuring topography of sample,the interaction between tip and sample will cause the deformation on the sample surface and the deformations will be different in the regions with different mechanical property.Furthermore,the deformations will cause non-negligible error in topography measurement.Theory analyses are carried out on this measurement error and further numerical research shows that mixed signals with high-frequency components are able to eliminate the measurement error.However,it is difficult to measure high-frequency signal for normal cantilever.In order to gain a proper high-frequency signal,a modified cantilever is introduced.The new designed cantilever can help us to adjust the relative values of eigen-frequencies of a cantilever and then enhance the high-frequency signal
来源 力学学报 ,2011,43(1):112-121 【核心库】
关键词 AFM ; 形貌测量偏差 ; 高频信号 ; 结构设计 ; 各阶本征频率相对值
地址

中国科学院力学研究所, 非线性力学国家重点实验室, 北京, 100190

语种 中文
文献类型 研究性论文
ISSN 0459-1879
学科 机械、仪表工业
基金 国家自然科学基金资助项目
文献收藏号 CSCD:4126884

参考文献 共 14 共1页

1.  Binnig G. Atomic force microscope. Physical Review Letters,1986,56(9):930-933 被引 463    
2.  Erlandsson R. equivalent atoms and imaging mechanisms in ac-mode atomic-force microscopy of Si(111)7x7. Physical Review B,1996,54(12):R8309-R8312 被引 2    
3.  Hitomi J. Atomic force microscopy in histology and cytology. Archives of histology and cytology,1996,59(5):421-431 被引 1    
4.  李喜德. 探针实验力学. 实验力学,2007,22(3):217-228 被引 9    
5.  李凯. 微梁传感研究大分子构象转变. 实验力学,2007,22(3):395-400 被引 2    
6.  Sahin O. Resonant harmonic response in tapping-mode atomic force microscopy. Physical Review B,2004,69(16):165416-1-165416-9 被引 5    
7.  Melikyan H. Hard disk magnetic domain nano-spatial resolution imaging by using a near-field scanning microwave microscope with an AFM probe tip. Journal of Magnetism and Magnetic Materials,2009,321:2483-2487 被引 1    
8.  Ohnesorge F. True atomic resolution by atomic force microscopy through repulsive and attractive forces. Science,1993,260(5113):1451-1456 被引 6    
9.  Liang J. Nanografting of alkanethiols by tapping mode atomic force microscopy. Langmuir,2007,23(11):6142-6147 被引 1    
10.  Solares SD. Frequency and force modulation atomic force microscopy: low-impact tapping-mode imaging without bistability. Measurement Science and Technology,2007,18(7):L9-L14 被引 1    
11.  Wang Y. Understanding tappingmode atomic force microscopy data on the surface of soft block copolymers. Surface Science,2003,530(3):136-148 被引 6    
12.  Yang CW. Imaging of soft matter with tapping-mode atomic force microscopy and noncantact-mode atomic force microscopy. Nanotechnology,2007,18(8):084009-1-084009-8 被引 3    
13.  Hertz J. On the contact of elastic solids. Journal fur die reine und angewandte Mathematik,1882,92:156-171 被引 1    
14.  Geradin M. Mechanical Vibrations Theory and Application to Structural Dynamics,1994 被引 1    
引证文献 2

1 魏征 轻敲模式下原子力显微镜的能量耗散 力学学报,2017,49(6):1301-1311
被引 7

2 胡璐 黏性流体环境下V型悬臂梁结构流固耦合振动特性研究 力学学报,2018,50(3):643-653
被引 12

显示所有2篇文献

论文科学数据集
PlumX Metrics
相关文献

 作者相关
 关键词相关
 参考文献相关

版权所有 ©2008 中国科学院文献情报中心 制作维护:中国科学院文献情报中心
地址:北京中关村北四环西路33号 邮政编码:100190 联系电话:(010)82627496 E-mail:cscd@mail.las.ac.cn 京ICP备05002861号-4 | 京公网安备11010802043238号