用于药品无损分析的近红外光谱检测系统光学设计
Optical design of near-infrared spectroscopy detection system for drug nondestructive analysis
查看参考文献12篇
文摘
|
为实现药品的无损快检,设计了基于傅里叶变换红外光谱检测的光学系统,系统工作波段为4000~12000 cm~(-1)。设计合适的光学结构并确定光学元件的特征参数,用Zemax对设计的光学系统进行光线追迹和优化分析。搭建实验平台,采集了5种相似药品的近红外光谱数据,用主成分分析法(PCA)结合误差反向传播人工神经网络算法(BPANN)对数据进行处理,有效区分了5种药品的光谱,识别准确率达到93.333%。实验结果表明:该近红外光谱检测系统适用于药品的无损分析,能有效鉴别相似药品。 |
其他语种文摘
|
To achieve fast nondestructive detection of drugs, the optical system of infrared spectroscopy detection based on Fourier transform is designed. The operating band of system ranges from 4000 cm~(-1) to 12000 cm~(-1). The suitable optical structure is designed and characteristic parameters of the optical components are determined. Ray tracing and optimization analysis on optical system designed are carried out with ZEMAX. The experimental platform is set up, and the near-infrared spectral data of five similar drugs are collected. The data are processed by combining principal component analysis (PCA) with error back propagation artificial neural network (BPANN). The spectra of five drugs are effectively identified and the recognition accuracy ratio reaches 93.333%. Experimental results show that the near-infrared spectroscopy detection system is suitable for non-destructive analysis of drugs, and can effectively identify similar drugs. |
来源
|
量子电子学报
,2017,34(3):264-271 【扩展库】
|
DOI
|
10.3969/j.issn.1007-5461.2017.03.002
|
关键词
|
光谱学
;
近红外光谱
;
药品分析
;
光学设计
;
傅里叶变换
|
地址
|
中国科学院安徽光学精密机械研究所, 中国科学院环境光学与技术重点实验室, 安徽, 合肥, 230031
|
语种
|
中文 |
文献类型
|
研究性论文 |
ISSN
|
1007-5461 |
学科
|
物理学 |
基金
|
国家重大科学仪器设备开发专项
|
文献收藏号
|
CSCD:6002784
|
参考文献 共
12
共1页
|
1.
Chen Jinzhui.
中国药品市场报告,2012
|
被引
1
次
|
|
|
|
2.
Chen Fenfen. Discriminant analysis for levocetirizine hydrochloride tablets based on near-infrared diffusion spectroscopy.
量子电子学报,2015,32(1):1-7
|
被引
1
次
|
|
|
|
3.
Liu Jiashui.
近红外光谱技术应用于南板蓝质量评价方法的研究. Master Thesis,2013
|
被引
1
次
|
|
|
|
4.
Karande A D. In-line quantification of micronized drug and excipients in tablets by near infrared (NIR) spectroscopy: Real time monitoring of tabletting process.
International Journal of Pharmaceutics,2010,396(1/2):63-74
|
被引
4
次
|
|
|
|
5.
Qu Ruyan. Application of near infrared spectrum in rapid identification of drugs.
实用医技杂志,2010,17(7):649-650
|
被引
1
次
|
|
|
|
6.
Heise H M. Mid-infrared attenuated total reflection spectroscopy of human stratum corneum using a silver halide fiber probe of square cross-section and adhesive tape stripping.
Powder Technology,2003,662:381-389
|
被引
1
次
|
|
|
|
7.
Scheibelhofer O. Spatially resolved monitoring of powder mixing processes via multiple NIR-probes.
Powder Technology,2013,243:161-170
|
被引
8
次
|
|
|
|
8.
Bruckner C. Optimal arrangement of 90 off-axis parabolic mirrors in THz setups.
Optik-International Journal for Light and Electron Optics,2010,121(1):113-119
|
被引
2
次
|
|
|
|
9.
Weng Shifu.
傅里叶变换红外光谱分析. 2nd Edition,2010
|
被引
1
次
|
|
|
|
10.
Yang Kun.
傅里叶变换红外光谱仪若干核心技术研究及其应用. Doctorial Dissertation,2010
|
被引
1
次
|
|
|
|
11.
Griffiths P R.
Fourier Transform Infrared Spectrometry. 2nd Edition,2007
|
被引
3
次
|
|
|
|
12.
Ma Bingxin.
时间调制型傅里叶光谱仪光学系统设计与杂光分析. Master Thesis,2012
|
被引
1
次
|
|
|
|
|