用于药品无损分析的近红外光谱检测系统光学设计
Optical design of near-infrared spectroscopy detection system for drug nondestructive analysis

查看参考文献12篇

文摘 为实现药品的无损快检,设计了基于傅里叶变换红外光谱检测的光学系统,系统工作波段为4000~12000 cm~(-1)。设计合适的光学结构并确定光学元件的特征参数,用Zemax对设计的光学系统进行光线追迹和优化分析。搭建实验平台,采集了5种相似药品的近红外光谱数据,用主成分分析法(PCA)结合误差反向传播人工神经网络算法(BPANN)对数据进行处理,有效区分了5种药品的光谱,识别准确率达到93.333%。实验结果表明:该近红外光谱检测系统适用于药品的无损分析,能有效鉴别相似药品。
其他语种文摘 To achieve fast nondestructive detection of drugs, the optical system of infrared spectroscopy detection based on Fourier transform is designed. The operating band of system ranges from 4000 cm~(-1) to 12000 cm~(-1). The suitable optical structure is designed and characteristic parameters of the optical components are determined. Ray tracing and optimization analysis on optical system designed are carried out with ZEMAX. The experimental platform is set up, and the near-infrared spectral data of five similar drugs are collected. The data are processed by combining principal component analysis (PCA) with error back propagation artificial neural network (BPANN). The spectra of five drugs are effectively identified and the recognition accuracy ratio reaches 93.333%. Experimental results show that the near-infrared spectroscopy detection system is suitable for non-destructive analysis of drugs, and can effectively identify similar drugs.
来源 量子电子学报 ,2017,34(3):264-271 【扩展库】
DOI 10.3969/j.issn.1007-5461.2017.03.002
关键词 光谱学 ; 近红外光谱 ; 药品分析 ; 光学设计 ; 傅里叶变换
地址

中国科学院安徽光学精密机械研究所, 中国科学院环境光学与技术重点实验室, 安徽, 合肥, 230031

语种 中文
文献类型 研究性论文
ISSN 1007-5461
学科 物理学
基金 国家重大科学仪器设备开发专项
文献收藏号 CSCD:6002784

参考文献 共 12 共1页

1.  Chen Jinzhui. 中国药品市场报告,2012 被引 1    
2.  Chen Fenfen. Discriminant analysis for levocetirizine hydrochloride tablets based on near-infrared diffusion spectroscopy. 量子电子学报,2015,32(1):1-7 被引 1    
3.  Liu Jiashui. 近红外光谱技术应用于南板蓝质量评价方法的研究. Master Thesis,2013 被引 1    
4.  Karande A D. In-line quantification of micronized drug and excipients in tablets by near infrared (NIR) spectroscopy: Real time monitoring of tabletting process. International Journal of Pharmaceutics,2010,396(1/2):63-74 被引 4    
5.  Qu Ruyan. Application of near infrared spectrum in rapid identification of drugs. 实用医技杂志,2010,17(7):649-650 被引 1    
6.  Heise H M. Mid-infrared attenuated total reflection spectroscopy of human stratum corneum using a silver halide fiber probe of square cross-section and adhesive tape stripping. Powder Technology,2003,662:381-389 被引 1    
7.  Scheibelhofer O. Spatially resolved monitoring of powder mixing processes via multiple NIR-probes. Powder Technology,2013,243:161-170 被引 8    
8.  Bruckner C. Optimal arrangement of 90 off-axis parabolic mirrors in THz setups. Optik-International Journal for Light and Electron Optics,2010,121(1):113-119 被引 2    
9.  Weng Shifu. 傅里叶变换红外光谱分析. 2nd Edition,2010 被引 1    
10.  Yang Kun. 傅里叶变换红外光谱仪若干核心技术研究及其应用. Doctorial Dissertation,2010 被引 1    
11.  Griffiths P R. Fourier Transform Infrared Spectrometry. 2nd Edition,2007 被引 3    
12.  Ma Bingxin. 时间调制型傅里叶光谱仪光学系统设计与杂光分析. Master Thesis,2012 被引 1    
引证文献 2

1 胡凯 太阳光谱地基跟踪遥感系统光学设计 量子电子学报,2021,38(3):290-300
被引 0 次

2 黄民德 基于移动平台的便携式FTIR灾害现场气体检测技术 量子电子学报,2024,41(2):235-245
被引 0 次

显示所有2篇文献

论文科学数据集
PlumX Metrics
相关文献

 作者相关
 关键词相关
 参考文献相关

版权所有 ©2008 中国科学院文献情报中心 制作维护:中国科学院文献情报中心
地址:北京中关村北四环西路33号 邮政编码:100190 联系电话:(010)82627496 E-mail:cscd@mail.las.ac.cn 京ICP备05002861号-4 | 京公网安备11010802043238号